Infant Equipment

MM-IC005 Transport Infant Incubator

Product Description

Introduction :

This unit is used for transporting the newborn babies between sickrooms or hospitals; AC Power and DC Power can be used alternatively and the device can be connected with Power 12V or 24V. Air-temp or skin-temp servo-controlled by micro-computer; double walls hood could prevent heat losing from hood and prevent the influence to the incubator from exterior heat resource; quite operation windows can reduce the disturbance to the baby during nursing with oxygen cylinder and oxygen supply system;; touch switch panel with LED digital display, set-temp, skin-temp, real-temp, storage battery voltage are displayed separately; indicated electricity-offer mode; night light, luminescence design; seven self-checking alarms with audible and visible; heating power rate indicates 0-100%; shockproof and adjustable stand with cylinder and three steps height.

Features :

  • Air mode and Baby mode controlled by micro-computer
  • Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance
  • Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately
  • Failure alarm indication
  • Double wall Acrylic glass hood with side door so that the bassinet can be pulled out from it
  • Natural air flow humidity
  • The height of whole unit can be adjusted
  • With Oxygen cylinder and Oxygen supply configuration
  • Observation lamp.

Standard configure:

  • Main body include (hood, controller, bassinet, internal battery and observation lamp)
  • Oxygen cylinders
  • Oxygen supply system
  • Infant bed
  • Skin temperature sensor
  • I.V. Pole
  • Mattress
  • Adjustable stand
  • Optional configuration : >370 override temperature mode

Technical Specification :

  • Power supply : AC220V~230V.50HZ/60 HZ
  • DC power : DC12V or DC24V
  • Power input : ≤400VA
  • Air Temperature control range.25℃~37℃
  • Skin Temperature control range.34℃~37℃
  • Skin sensor precision : ≤0.3℃
  • Temperature variability : ≤1.0℃
  • Uniformity of bed temperature: ≤1.5℃ (bed is on level position)
  • Continue working time (environmental temperature 15℃.control temperature 36℃).90min(1pc of storage battery)
  • Noise inside hood : ≤55dB (A), Environment noise: ≤45dB

Alarming :

  • Over-temp alarm
  • Temperature Deviation alarm
  • Motor fan failed alarm
  • Sensor alarm
  • Power failure alarm